FDA recall record

TomoTherapy Treatment System with software versions 2.0.1 / 2.0.2 / 2.0.3/ recall — FDA Z-0507-2015

Accuray has become aware of a potential safety issue related to the TomoTherapy Treatment System caused by a failure to monitor the jaw position after a…

Class IIclassification
Z-0507-2015official record ID
Dec 17, 2014report date

Recall record at a glance

Official recordZ-0507-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmTomoTherapy Inc.
Report dateDec 17, 2014
Recall initiation dateNov 21, 2014
Quantity210 ( 72 USA; 138 OUS )
LocationMadison, WI, United States

Official product description

TomoTherapy Treatment System with software versions 2.0.1 / 2.0.2 / 2.0.3/ (Hi:Art 5.0.1 / 5.0.2 / 5.0.3)

Why the product was recalled

Accuray has become aware of a potential safety issue related to the TomoTherapy Treatment System caused by a failure to monitor the jaw position after a jaw error occurs. This may result in an incorrect jaw position during treatment, without generating system interruption.

Distribution information

Worldwide Distribution - USA (nationwide) included the states of : AL, AZ, AR, CA, FL, GA, IL, IN, KS, ,Y, MD, MN, MO, NE, NM, NY, NC, OK, OR, PA, SD, TN, TX, UT, WA, WV, WI and WY., and the countries of : JAPAN, BELGIUM, NETHERLANDS, ITALY, GERMANY, KOREA, FRANCE, SWITZERLAND, ENGLAND, CANADA, MEXICO, HONG KONG, AUSTRALIA, SRI LANKA, GREECE, CHINA, PHILIPPINES, MARTINIQUE, ROMANIA, SPAIN and TURKEY.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.