Recall record at a glance
| Official record | Z-2063-2018 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Orthofix Srl |
| Report date | Jun 13, 2018 |
| Recall initiation date | Jan 30, 2018 |
| Quantity | 408 units |
| Location | Bussolengo, N/A, Italy |
Official product description
TL-HEX TruLok Hexapod System, Double Row Footplates, TL-HEX Sterile, All Sizes: 120 MM, 140 MM, 160 MM, 180 MM, 200 MM. The TL-HEX TrueLock Hexapod System (TL-HEX) consists of rings, foot plates and struts to be used in conjunction with the TruLok external fixation system. The system is a metal bone fixation device for stabilization and correction during the normal healing process. The purpose of the TL-HEX System is to provide a solution for deformity correction and keeping the bone segments aligned during the healing process. The device may only be applied by a physician, who is fully responsible for the surgical and post-operative procedures of the system and is able to manage possible mechanical and software limitations. Indications for use for adults and all pediatric subgroups except newborns include: post-traumatic joint contracture which has resulted in loss of range of motion, fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction, open and closed fracture fixation, pseudoarthrosis of long bones, limb lengthening by epiphyseal or metaphyseal distraction, correction of bony or soft tissue deformities, correction of bony of soft tissue defects, joint arthrodesis, infected fractures or non unions.
Why the product was recalled
The firm received complaints regarding the incompatibility and dimensional interference between the double row footplates manufactured in a limited period of 2014 and struts manufactured after September 2016. The firm determined that certain batches of the double row footplates are not compatible with the current version of the TL-HEX struts.
Distribution information
Worldwide Distribution - US Distribution within Texas., and to the countries of : Australia, Belgium, Chile, Denmark, France, Germany, Israel, Italy, Japan, South Africa, Spain, Sweden, Switzerland and United Kingdom.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.