Recall record at a glance
| Official record | Z-1632-2018 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class III |
| FDA record status | Terminated |
| Recalling firm | Michigan Instruments, Inc. |
| Report date | May 16, 2018 |
| Recall initiation date | Feb 13, 2018 |
| Quantity | 38 devices |
| Location | Grand Rapids, MI, United States |
Official product description
Thumper Cardiopulmonary Resuscitator, Model 1007CCV, Catalog No. 15370 Product Usage: The device is used to perform Cardiopulmonary Resuscitation (CRP) on adult patients and only adult patients in cases of clinical death as defined by a lack of spontaneous breathing and pulse.
Why the product was recalled
A performance issue with the chest depth needle valve (CDNV) may arise where turning the knob will not deliver any compression depth to the piston during the first 5 revolutions of the knob.The last ¿ to ¿ revolution of the knob could deliver a full 8cm of compression.
Distribution information
Distributed to China.
How to read the classification
FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.