FDA recall record

Thoratec CentriMag Motor, Model: 102956, a component of the Thoratec CentriMag Blood Pumping recall — FDA Z-0103-2019

Reports of circulatory support system motor stopping. Motor stop can be attributed to a break of a wire or short circuit between wires within the…

Class IIclassification
Z-0103-2019official record ID
Oct 24, 2018report date

Recall record at a glance

Official recordZ-0103-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmThoratec Switzerland GMBH
Report dateOct 24, 2018
Recall initiation dateSep 5, 2018
Quantity1569
LocationZurich, N/A, Switzerland

Official product description

Thoratec CentriMag Motor, Model: 102956, a component of the Thoratec CentriMag Blood Pumping System and PediMag Blood Pumping System.

Why the product was recalled

Reports of circulatory support system motor stopping. Motor stop can be attributed to a break of a wire or short circuit between wires within the electrical cable where the cable exits the motor cable bend protection.

Distribution information

U.S.: OR, AZ, CA, ID, WA, WI, NY, MA, CT, NJ, ME, RI, IL, MI, TN, OH, IN, KY, IA, AL, FL, NC, GA, SC, VA, MD, DC, PA, WV, DE, MN, MT, NE, CO, NM, UT, AR, MO, MC, TX, OK, LA, MS, ND

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.