Official product description
There is a complaint regarding internal check cable guide mechanism failure for the Nucletron Flexitron brachytherapy afterloader system. Due to an obstruction during the check cable out drive, the internal guiding tube became displaced and the check cable became damaged. This issue could cause treatment interruption or in a very unlikely scenario, incorrect source positioning and radiation treatment errors.
Why the product was recalled
A component failure may lead to treatment interruption or incorrect source positioning.
Distribution information
Distribution in United States and worldwide
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.