FDA recall record

Beckman Coulter The UniCel DxI 600 and 800 Access Immunoassay Systems, that are connected to recall — FDA Z-1077-2019

When racks are front-loaded onto the sample presentation unit while samples from the automation line are being processed, a scheduling error within the…

Class IIclassification
Z-1077-2019official record ID
Apr 10, 2019report date

Recall record at a glance

Official recordZ-1077-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmBeckman Coulter Inc.
Report dateApr 10, 2019
Recall initiation dateDec 5, 2018
Quantity2842
LocationBrea, CA, United States

Official product description

The UniCel DxI 600 and 800 Access Immunoassay Systems, that are connected to an automation line. UniCel DxI 600 Catalog Numbers: A30260, A71460, A71461, A92060. Unicel DxI 800 Catalog Numbers: 973100, A71456, A71457, A84545, A25288, A25285.

Why the product was recalled

When racks are front-loaded onto the sample presentation unit while samples from the automation line are being processed, a scheduling error within the software can cause mishandling of samples during the creation of an aliquot and lead to incorrect results being generated. Affected samples typically yield non-reproducible falsely decreased results by up to 100%, however the potential for falsely elevated results cannot be ruled out.

Distribution information

US: NJ, OH, NY, GA, FL, TN, WI, MI, MS, CA, NC, TX, PA, DC, IN, KY, IL, WA, MN, LA, ND, ID, WV, AK, AL, MA. OUS: Argentina, Australia, Bahrain, Canada, China, Croatia, Czech Republic, Egypt, Germany, Greece, India, Italy, Malaysia, Morocco, Netherlands, Nigeria, Poland, Portugal, Qatar, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Thailand, Tunisia, United Kingdom, United States, Vietnam

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.