FDA recall record

Schaerer Medical USA The surgical table is a stand alone unit recall — FDA Z-0371-2017

The floor locking cylinders contained in the base of the Schaerer 7300 model surgical tables were manufactured with the incorrect "O" ring which may cause…

Class IIclassification
Z-0371-2017official record ID
Nov 16, 2016report date

Recall record at a glance

Official recordZ-0371-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSchaerer Medical USA Inc.
Report dateNov 16, 2016
Recall initiation dateSep 27, 2016
Quantity4 surgical tables with defective floor locking cylinders were distributed. Each table contains 3 cylinders for a total of 12 cylinders.
LocationCincinnati, OH, United States

Official product description

The surgical table is a stand alone unit. Three (3) floor locking cylinders are attached to the base of the surgical table. The individual cylinders are replaceable. Product Usage: The cylinders are used to lock the surgical table in place.

Why the product was recalled

The floor locking cylinders contained in the base of the Schaerer 7300 model surgical tables were manufactured with the incorrect "O" ring which may cause the cylinder to leak.

Distribution information

Worldwide Distribution - US Nationwide in the states of: CA, CT, FL, IA, IN, MI, MN, MO, OH, PA, and VA Replcement cylinders wre distributed

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.