FDA recall record

Sophysa S.A The Sophysa Pressio Ventricular Intracranial Pressure Monitoring Kit monitoring recall — FDA Z-1826-2024

An increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected…

Class IIclassification
Z-1826-2024official record ID
May 22, 2024report date

Recall record at a glance

Official recordZ-1826-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmSophysa S.A.
Report dateMay 22, 2024
Recall initiation dateApr 9, 2024
Quantity3175 units- 41 units US; 3134 units OUS
LocationOrsay, N/A, France

Official product description

The Sophysa Pressio Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial ventricular pressure and temperature with tunneling (PSO-VTT) indicated for use in intraventricular pressure monitoring and cerebrospinal fluid drainage application. Model/Catalog Number: PSO-VTT

Why the product was recalled

An increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected during implantation, it could pose an infectious risk to the patient. The cause of the leaks is indicated to be manufacturing error involving perforation of the inner channel of the catheter during stylet insertion.

Distribution information

Worldwide - US Nationwide distribution in the states of AZ, NY, TX and the countries of ARGENTINA, CHINA, CZECH REPUBLIC, FRANCE, GREECE, HUNGARY, JAPAN, PAKISTAN, Portugal, ROMANIA, SOUTH AFRICA, SWEDEN, VIETNAM.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.