FDA recall record

Physio-Control The LIFEPAK 15 monitor/defibrillator is intended for use by trained medical recall — FDA Z-0761-2017

Customer complaint was received relating to a LifePak15 device that would power on then power down unexpectedly after a short period of use. When the LP15…

Class IIclassification
Z-0761-2017official record ID
Dec 21, 2016report date

Recall record at a glance

Official recordZ-0761-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmPhysio-Control, Inc.
Report dateDec 21, 2016
Recall initiation dateMar 11, 2010
Quantity1399 units (866 in the US and 533 outside US)
LocationRedmond, WA, United States

Official product description

The LIFEPAK 15 monitor/defibrillator is intended for use by trained medical personnel in outdoor and indoor emergency care settings within the environmental conditions specified. Product Part Numbers V-15-2-000XXX

Why the product was recalled

Customer complaint was received relating to a LifePak15 device that would power on then power down unexpectedly after a short period of use. When the LP15 powered down, power could be immediately restored by activation of the power button. The device was connected to a cellular modem from MultiTech systems for use with LIFE NET services. The corrective action was completed by Velocitor Solutions in December 2010.

Distribution information

worldwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records