FDA recall record

Stryker Sustainability Solutions The Fuhrman Pleural/Pneumopericardial Drainage Set The recall — FDA Z-0478-2016

Stryker Sustainability Solutions is recalling the Fuhrman Pleural & Pneumopericardial Drainage Set due to receiving complaints that it broke off in the…

Class Iclassification
Z-0478-2016official record ID
Jan 13, 2016report date

Recall record at a glance

Official recordZ-0478-2016
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmStryker Sustainability Solutions
Report dateJan 13, 2016
Recall initiation dateNov 17, 2015
Quantity34 devices
LocationPhoenix, AZ, United States

Official product description

The Fuhrman Pleural/Pneumopericardial Drainage Set The Fuhrman Pleural/Pneumopericardial Drainage Set consists of pigtail catheter, access needle, dilator, and wire guide. The set also includes a multipurpose tubing adapter, and three-way stopcock.

Why the product was recalled

Stryker Sustainability Solutions is recalling the Fuhrman Pleural & Pneumopericardial Drainage Set due to receiving complaints that it broke off in the pleural cavity while inserting the device into the patient.

Distribution information

US Distribution in the states of: AK, CA, WI, IN, FL, OR, NY, and NC.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records