FDA recall record

SIE AG, Surgical Instrument Engineering The FEMTO LDV(TM)Z8 Femtosecond Surgical Laser The recall — FDA Z-1272-2018

This Correction is being initiated based on a customer complaint received on November 16, 2017 which indicated a complication with a cataract procedure.…

Class IIclassification
Z-1272-2018official record ID
Apr 4, 2018report date

Recall record at a glance

Official recordZ-1272-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSIE AG, Surgical Instrument Engineering
Report dateApr 4, 2018
Recall initiation dateJan 21, 2018
Quantity125 devices
LocationPort, Switzerland

Official product description

The FEMTO LDV(TM)Z8 Femtosecond Surgical Laser The FEMTO LDV(TM)Z8 Femtosecond Surgical Laser is an ophthalmic surgical laser intended for use in the creation of corneal incisions indicated for use in patients undergoing LASIK surgery, tunnel creation for implantation of rings, pocket creation for implantation of corneal implants, lamellar keratoplasty, penetrating keratoplasty or other treatment requiring lamellar resection of the cornea at a varying depth with respect to the corneal surface. In addition, the FEMTO LDV(TM)Z8 Surgical Laser is intended for use in the creation of capsulotomy, phacofragmentation and the creation of single plane, multi-plane, arc cuts/incisions in the cornea, each of which may be performed either individually or consecutively during the same procedure indicated for use in patients undergoing cataract surgery for removal of the crystalline lens.

Why the product was recalled

This Correction is being initiated based on a customer complaint received on November 16, 2017 which indicated a complication with a cataract procedure. An investigation completed by the manufacturer determined that although the complication was due to user error improvements could and should be made on our part to improve the usability and performance of our device.

Distribution information

Worldwide Distribution: US (nationwide) including states of: AL, CA, FL, GA, IL, IN, MN, OR, TN, TX, VA, and WA; and countries of: Argentina, Austria, Belgium, Brazil, Canada, France, Germany, India, Iraq, Israel, Italy, Japan, Lebanon, Malaysia, Mexico, Namibia, Pakistan, Peru, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Spain, South Africa, Switzerland, Taiwan, and United Arab Emirates.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Telephone.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.