FDA recall record

EKOS The EKOS EkoSonic Control Unit is intended exclusively for use with the EkoSonic recall — FDA Z-2148-2015

The Connector Interface Cable (CIC) was not recognized by the EkoSonic PT-3B control unit.

Class IIclassification
Z-2148-2015official record ID
Jul 29, 2015report date

Recall record at a glance

Official recordZ-2148-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmEKOS Corporation
Report dateJul 29, 2015
Recall initiation dateJun 24, 2015
Quantity46 units
LocationBothell, WA, United States

Official product description

The EKOS EkoSonic Control Unit is intended exclusively for use with the EkoSonic Endovascular Device and the EkoSonic MACH4 Endovascular Device. The EkoSonic Endovascular System consists of three main components: -a single use EkoSonic Endovascular Device which is comprised of an Intelligent Drug Delivery Catheter (IDDC) and -removable MicroSonic Device (MSD), and -a reusable EKOS EkoSonic Control System. During use, fluids are delivered through the side-holes of the IDDC. The MSD is placed through the central lumen of the IDDC to deliver ultrasound energy along the length of fluid delivery, dispersing the fluid into the intravascular treatment site. The reusable EKOS EkoSonic Control Unit provides power to the system and the user interface for operator control. A reusable, nonsterile Connector Interface Cable connects the EKOS EkoSonic Control Unit to the EkoSonic Endovascular Device.

Why the product was recalled

The Connector Interface Cable (CIC) was not recognized by the EkoSonic PT-3B control unit.

Distribution information

Nationwide Distribution

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceSep 28, 2016

EKOS recall — FDA Z-2854-2016

The recalling firm identified ten devices which were distributed and could potentially be non-compliant. These devices potentially underwent a routine manufacturing test with non-compliant…

Class IIFDA: TerminatedEKOS Corporation
Medical DeviceMar 2, 2016

EKOS recall — FDA Z-0864-2016

The EkoSonic kit, including the individually labeled MicroSonic Device (MSD) and Intelligent Drug Delivery Catheter (IDDC) are labeled with a 135 cm working length, and programmed with a…

Class IIFDA: TerminatedEKOS Corporation