Ra Medical Systems The DABRA Laser (model RA-308) is designed to provide ultraviolet laser recall — FDA Z-2218-2019

Lasers/Catheters did not calibrate during set-up prior to use.

Class IIclassification
Z-2218-2019official record ID
Aug 14, 2019report date

Recall record at a glance

Official recordZ-2218-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmRa Medical Systems Inc.
Report dateAug 14, 2019
Recall initiation dateFeb 15, 2018
Quantity4
LocationCarlsbad, CA, United States

Official product description

The DABRA Laser (model RA-308) is designed to provide ultraviolet laser energy endovascularly through a catheter to occluded lesions in the peripheral arteries to cross them. The system is a self-contained unit that consists of a light source unit and a single use catheter. The light source contains a microprocessor with software to provide safety features and a user-friendly interface. The laser operates by producing a high current discharge within a chamber that is filled with rare gases and halogen. The discharge produces pulses of UV light, which is amplified by reflecting the energy within the resonating chamber many times. The output is directed to the desired target location by the catheter. A footswitch connected to the light source controls the laser action.

Why the product was recalled

Lasers/Catheters did not calibrate during set-up prior to use.

Distribution information

U.S: CA, PA

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.