Recall record at a glance
| Official record | Z-0865-2014 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class III |
| FDA record status | Terminated |
| Recalling firm | GE Healthcare It |
| Report date | Feb 5, 2014 |
| Recall initiation date | Nov 20, 2012 |
| Quantity | 7 units |
| Location | Barrington, IL, United States |
Official product description
The Centricity Laboratory System is intended to be an information system designed to support the clinical and administrative activities associated with the provision and utilization of clinical laboratory services and facilities, e.g., the storing and delivering of analytical results. It is a specially designed data program application (software), which is supplied for installation in existing mainframe or decentralized computers or a computer network. This software system will not be the software directly used to run, steer, or control any specific laboratory analyzer or equipment. It is intended for use in laboratory facilities, including a central laboratory or in a multiple laboratory environment servicing satellite laboratories, reference laboratories, clinics, etc. Centricity Laboratory System is not intended for direct patient contact. The Centricity Laboratory System product is designed to facilitate the general clinical, anatomic pathology and cytology laboratory workflow such as order entry, results entry, instrument interfacing, results reporting and patient record retrieval. It integrates with other Hospital Information Systems (HIS) through HL7 interface by order entering and also report processing. The Centricity Laboratory System is intended to interface with various lab results. Centricity Laboratory System is intended to be used by laboratory technologists, technicians, analyst and other trained/site authorized system users
Why the product was recalled
GE Healthcare is aware of a potential safety issue associated with the use of GE Centricity Laboratory Core Lab calculated results in two scenarios. In Scenario 1: Calculated results greater than 6 digits plus a decimal are reported as 0 instead of TOO BIG when significant figures are not defined for the item. In Scenario 2: Calculated results greater than 6 digits plus a decimal are trunc
Distribution information
Worldwide Distribution - US (Nationwide) to states of: MN, GA, and CA; and International to: England.
How to read the classification
FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.