FDA recall record

Integra Life Sci The 2.7mm Drilling Guide is a reusable instrument used in conjunction with recall — FDA Z-2381-2012

Integra' s quality system noted an adverse trend in complaints for breakage of the drilling guides during use. None of the complaints resulted in patient…

Class IIclassification
Z-2381-2012official record ID
Sep 26, 2012report date

Recall record at a glance

Official recordZ-2381-2012
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmIntegra Life Sci.
Report dateSep 26, 2012
Recall initiation dateJun 14, 2012
Quantity2,389 units, Catalogue No. 219635ND; 2.7mm Drilling Guides
LocationPlainsboro, NJ, United States

Official product description

The 2.7mm Drilling Guide is a reusable instrument used in conjunction with a drill to prepare oriented holes in bone for insertion of a Standard Surfix Locking Screw with plates for the Tiblaxis, Uni-CP, Advansys MLP and Advansys DLP compression plate systems. The Drilling Guides are initially supplied to users in kits containing the devices that comprise the Tibiaxis, Uni-CP, Advansys MLP and Advansys DLP systems. Should the Drilling Guides need to be replaced, they can be replenished by the customer placing an order. The replacement is shipped to the customer in sealed plastic bags. Uni-CP" Compression Plate The Uni-CP" Compression Plate is indicated for fixation of bone fractures or for bone reconstruction: - Arthrodesis in foot and ankle surgery - Fracture management in the foot and ankle - Mono or bi-cortical osteotomies in the foot and ankle TIBIAXYS" Plates The Newdeal TIBIAXYS" Plates are indicated For fixation of bone fractures or for bone reconstruction. Examples include Osteotomies and fractures of ankle joint, distal tibia and fibula. The Newdeal TIBIAXYS" Plates must be fixed with the Newdeal Surfix 3.5mm diameter fixation screws and lock-screws. ADVANSYS Mid-Foot Plating System: Dorsal Lisfranc Plate (DLP) and Medical Lisfranc Plate (MLP) The Dorsal Lisfranc Plates are intended for fractures, fusions, osteotomies and replantations of small bones at the tarsometatarsal joints The Medial Lisfranc Plates are intended for bone fixation such as: arthrodesis of the 1st metatarsocuneiform joint to reposition and stabilize a metatarsus primus varus: - Lisfranc arthrodesis - Mono or bi-cortical osteotomies or fractures near the 1st metatarsocuneiform joint

Why the product was recalled

Integra' s quality system noted an adverse trend in complaints for breakage of the drilling guides during use. None of the complaints resulted in patient injuries but in some cases resulted in an increase in the time to complete the surgical procedure. A physician was consulted to assess patient risk and concluded that any potential for risk would be transient, self-limiting or minor, or result

Distribution information

Worldwide Distribution-USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NV, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WI and the countries of Austria, Czech republic, Denmark, France, Germany, Ireland, Israel, Italy, Kingdom of Saudi Arabia, Lebanon, Norway, Portugal, Switzerland, South Africa, Spain, Sweden, and United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: not stated.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.