Recall record at a glance
| Official record | Z-0157-2015 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Terumo Cardiovascular Systems Corp |
| Report date | Nov 12, 2014 |
| Recall initiation date | Mar 1, 2011 |
| Quantity | 165 units |
| Location | Ashland, MA, United States |
Official product description
Terumo Custom Integrated Cardioplegia Delivery Set-LINE FOR 15501/16015/5852 Item Number: 16110 Product Usage: This is a replacement line to multiple Cardioplegia Delivery Sets - and includes the female luer connector that is the subject of this correction/removal activity
Why the product was recalled
Defect in the luer connector can compromise the seal on the female side of the connector. and the compromised seal can introduce air in the table line
Distribution information
Worldwide Distribution: US (nationwide) and the countries of Canada, Brazil, Hong Kong, Korea, and Thailand.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.