FDA recall record

Remote Diagnostic Technologies Tempus LS - Manual Defibrillator recall — FDA Z-0564-2021

A software error was detected within software version 1.3.4 for the Tempus LS device. When exiting pacer back into manual mode the pacer pulse wave may…

Class IIclassification
Z-0564-2021official record ID
Dec 23, 2020report date

Recall record at a glance

Official recordZ-0564-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmRemote Diagnostic Technologies Ltd.
Report dateDec 23, 2020
Recall initiation dateNov 17, 2020
Quantity175 devices
LocationBasingstoke, N/A, United Kingdom

Official product description

Tempus LS - Manual Defibrillator, Model Number: 00-3020 Software Version: 1.3.4 - Product Usage: The device is intended to be used in the following environmental: - Primarily in pre-hospital care and transport applications (including ambulances, fixed and rotary wing aircraft).

Why the product was recalled

A software error was detected within software version 1.3.4 for the Tempus LS device. When exiting pacer back into manual mode the pacer pulse wave may put the ECG calibration out of calibration. This software error may result in a delay in therapy, as the clinician is not provided an ECG waveform on which a shockable rhythm decision can be made. The device is still able to deliver a manual defibrillation shock.

Distribution information

US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.