FDA recall record

Technidata S.A TDWorkstation version 11.01.A or higher recall — FDA Z-1310-2014

If a conflict occurs between a Demography Query and a Result Transmission by the instrument, the TDWorkstation could send an incorrect patient demography…

Class IIIclassification
Z-1310-2014official record ID
Apr 9, 2014report date

Recall record at a glance

Official recordZ-1310-2014
CategoryMedical Device Recalls
ClassificationClass III
FDA record statusTerminated
Recalling firmTechnidata S.A.
Report dateApr 9, 2014
Recall initiation dateFeb 17, 2014
Quantity85 units total (1 unit in US)
LocationMontbonnot St Martin, N/A, France

Official product description

TDWorkstation version 11.01.A or higher. TDWorkstation Evolution. TDWorkstation receives patient results from a connected instrument and sends them to a Laboratory Information System (LIS).

Why the product was recalled

If a conflict occurs between a Demography Query and a Result Transmission by the instrument, the TDWorkstation could send an incorrect patient demography to an instrument.

Distribution information

Distributed in the state of PA.

How to read the classification

FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.