FDA recall record

Siemens Healthcare Diagnostics Syva Emit Methotrexate Assay, is a homogeneous enzyme recall — FDA Z-1618-2016

Siemens Healthcare Diagnostics confirmed that Syva Emit Methotrexate Application (DVMTX.1) for Dimension Vista System does not meet On-Board Stability…

Class IIclassification
Z-1618-2016official record ID
May 18, 2016report date

Recall record at a glance

Official recordZ-1618-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSiemens Healthcare Diagnostics, Inc.
Report dateMay 18, 2016
Recall initiation dateMar 8, 2016
Quantity18364 units
LocationNewark, DE, United States

Official product description

Syva Emit Methotrexate Assay, is a homogeneous enzyme immunoassay intended for use in the quantitative analysis of methotrexate in human serum or plasma.

Why the product was recalled

Siemens Healthcare Diagnostics confirmed that Syva Emit Methotrexate Application (DVMTX.1) for Dimension Vista System does not meet On-Board Stability claim of 20 hours. Testing using reagent at the end of the 20 hours On-Board Stability has shown that values of methotrexate for patient and quality control samples increased.

Distribution information

Units were shipped to MI, FL, NJ, NY, DC, CA, GA, MI, MT, VA, NC, TX, PA, Australia, Belgium, Czech Republic, Finland, Japan, Republic of Korea, Italy, Saudi Arabia, Slovakia, Slovenia, Spain and Switzerland.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: N/A.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.