FDA recall record

GE Healthcare System, Tomography Computed Emmission recall — FDA Z-2259-2020

GE Healthcare has identified that a small number of Nuclear Medicine systems are operating without currently updated software. In a recent event, a…

Class IIclassification
Z-2259-2020official record ID
Jun 10, 2020report date

Recall record at a glance

Official recordZ-2259-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmGE Healthcare, LLC
Report dateJun 10, 2020
Recall initiation dateNov 30, 2018
Quantity526 total units
LocationWaukesha, WI, United States

Official product description

System, Tomography Computed Emmission

Why the product was recalled

GE Healthcare has identified that a small number of Nuclear Medicine systems are operating without currently updated software. In a recent event, a patient s finger was injured during the Unload process while the automatic Unload feature was being used; the software on the system was not current.

Distribution information

Domestic Distribution: AL, AR, AZ, CA, CO, CT, DC, DE, FL,GA, IA, IL, IN, KY, LA,MA, MD, ME, MI, MN, MO, MS, NC, NE, NM, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT,VA, WI and WV. Foreign Distribution: Argentina, Australia, Belgium, Brazil, Canada, Colombia, Denmark, France, Germany, Ghana, India, Israel, Italy, Japan, Korea, Malaisya, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Singapore, Sweden, Taiwan, Thailand, Turkey and United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.