Recall record at a glance
| Official record | Z-2259-2020 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | GE Healthcare, LLC |
| Report date | Jun 10, 2020 |
| Recall initiation date | Nov 30, 2018 |
| Quantity | 526 total units |
| Location | Waukesha, WI, United States |
Official product description
System, Tomography Computed Emmission
Why the product was recalled
GE Healthcare has identified that a small number of Nuclear Medicine systems are operating without currently updated software. In a recent event, a patient s finger was injured during the Unload process while the automatic Unload feature was being used; the software on the system was not current.
Distribution information
Domestic Distribution: AL, AR, AZ, CA, CO, CT, DC, DE, FL,GA, IA, IL, IN, KY, LA,MA, MD, ME, MI, MN, MO, MS, NC, NE, NM, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT,VA, WI and WV. Foreign Distribution: Argentina, Australia, Belgium, Brazil, Canada, Colombia, Denmark, France, Germany, Ghana, India, Israel, Italy, Japan, Korea, Malaisya, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Singapore, Sweden, Taiwan, Thailand, Turkey and United Kingdom.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.