FDA recall record

Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number recall — FDA Z-1113-2025

Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on…

Class IIclassification
Z-1113-2025official record ID
Feb 12, 2025report date

Recall record at a glance

Official recordZ-1113-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmSysmex America, Inc.
Report dateFeb 12, 2025
Recall initiation dateJan 6, 2025
Quantity32 units
LocationLincolnshire, IL, United States

Official product description

Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341. Product usage for automated pipetting, diluting, and specimen processing workstations for flow cytometric analysis.

Why the product was recalled

Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.

Distribution information

US Nationwide.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceApr 22, 2020

Sysmex The CF-70 instrument recall — FDA Z-1703-2020

Software mismatch-When the software versions between the SP-50 and CF-70 are not matched, and an error condition occurs requiring the operator to open the cover door of the CF-70 to remove…

Class IIFDA: TerminatedSysmex America, Inc.