FDA recall record

Siemens Medical Solutions USA syngo Application software VE20 (Material Number 10848815) recall — FDA Z-1339-2022

After CT image data from Toshiba is loaded, image mirroring can occur along the horizontal and vertical image axes. If this error occurs, the patient…

Class IIclassification
Z-1339-2022official record ID
Jul 13, 2022report date

Recall record at a glance

Official recordZ-1339-2022
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmSiemens Medical Solutions USA, Inc.
Report dateJul 13, 2022
Recall initiation dateFeb 25, 2022
Quantity127 units
LocationMalvern, PA, United States

Official product description

syngo Application software VE20 (Material Number 10848815) installed on the Artis pheno and Artis icono systems as follows: (1) Artis pheno Model Number:10849000 (2) Artis icono biplane Model Number: 11327600 (3) Artis icono floor Model Number: 11327700

Why the product was recalled

After CT image data from Toshiba is loaded, image mirroring can occur along the horizontal and vertical image axes. If this error occurs, the patient orientation/position may be misinterpreted and result in inappropriate treatment, even if the incorrect visualization is obvious.

Distribution information

US Nationwide.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records