FDA recall record

Hospira Symbiq One Channel Infuser recall — FDA Z-0305-2017

Symbiq Infusers have the potential to experience a white screen during titration of a Keep Vein Open (KVO) delivery. This can allow the clinician to…

Class IIclassification
Z-0305-2017official record ID
Nov 2, 2016report date

Recall record at a glance

Official recordZ-0305-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmHospira Inc.
Report dateNov 2, 2016
Recall initiation dateFeb 4, 2013
Quantity35,596 devices in total
LocationLake Forest, IL, United States

Official product description

Symbiq One Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids Device is a general purpose infusion pump designed to deliver fluids, solutions, medications, agents, nutritionals, electrolytes, blood and blood products for parenteral administration; and various solution through enteral, intravenous, intra-arterial, subcutaneous, and pathways.

Why the product was recalled

Symbiq Infusers have the potential to experience a white screen during titration of a Keep Vein Open (KVO) delivery. This can allow the clinician to select the "NEXT" button prior to completion of the programming. If this error occurs, a delay in therapy could occur and potentially result in significant injury.

Distribution information

Worldwide Distribution - USA (nationwide) and Canada

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceNov 16, 2016

Hospira Symbiq Two Channel Infuser recall — FDA Z-0362-2017

May go to a white screen due to a problem with the SOM2 module. A white screen is a diagnostic screen displayed to the user as part of the safety shutdown procedure of the device. The…

Class IIFDA: TerminatedHospira Inc.
Medical DeviceNov 16, 2016

Hospira Symbiq One Channel Infuser recall — FDA Z-0361-2017

May go to a white screen due to a problem with the SOM2 module. A white screen is a diagnostic screen displayed to the user as part of the safety shutdown procedure of the device. The…

Class IIFDA: TerminatedHospira Inc.
Medical DeviceNov 9, 2016

Hospira Symbiq Two Channel Infuser recall — FDA Z-0354-2017

Hospira has received reports of Malfunction S205 Backup Battery Failure alarms in Symbiq v3.13 pumps. This issue to an incorrect installation of components on some of the Power Supply…

Class IIFDA: TerminatedHospira Inc.