JK Products & Services SUN|DASH RADIUS 252 tanning system recall — FDA Z-2334-2012

The firm has tracked a systematic failure in the door cord on the Sun|Dash Radius 252 product line. In some instances when the door cord is installed with…

Class IIclassification
Z-2334-2012official record ID
Sep 19, 2012report date

Recall record at a glance

Official recordZ-2334-2012
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmJK Products & Services, Inc.
Report dateSep 19, 2012
Recall initiation dateJul 8, 2005
Quantity1948 units
LocationJonesboro, AR, United States

Official product description

SUN|DASH RADIUS 252 tanning system; 52 lamps, Ultraviolet Radiation. The device is intended to be used for tanning of the human skin.

Why the product was recalled

The firm has tracked a systematic failure in the door cord on the Sun|Dash Radius 252 product line. In some instances when the door cord is installed with little or no slack outside the upper decor piece, the cord has been known to fail causing a short in the wiring at the pivot point in the cord.

Distribution information

Worldwide Distribution - USA (nationwide) and Canada

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.