Official product description
Stryker, REF: 0250070460, 5.0 mm x 32 cm, Strykeprobe, Outer Sheath Replacement, RX Only, Non-Sterile, CE, UDI: (01) 07613327051971 These products are distributed with the following sheath/probe: 0250070441-PKG, StrykeFlow Electrocautery Probe, Spatula Tip 5MM 0250070442-PKG, StrykeFlow Electrocautery Probe, J Tip, 5MM 0250070443-PKG, StrykeFlow Electrocautery Probe, L Tip, 5MM 0250070444-PKG, StrykeFlow Electrocautery Probe, Ball Tip, 5MM 0250070445-PKG, StrykeFlow Electrocautery Probe, Needle Tip 5MM 0250070446-PKG, StrykeFlow Electrocautery Probe, Spoon Tip, 5MM
Why the product was recalled
Potential diff. in the length of the sheath tube and the sheath/probe. A diff. in length could lead to the base of the sheath/probe not being fully seated in the sheath tube which could result in the distal tip of the sheath melting. This could lead to a compromised insulation, leading to unintentional flow of electricity to the patient. The sheath/probe is packaged separately.
Distribution information
OUS: Netherlands, Spain, Switzerland, Colombia, Canada, Japan, Taiwan,, Puerto Rico, South Africa, and South Korea
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.