FDA recall record

Stryker Orthopaedics Specialty Triathlon Tibial Alignment Handle with Secondary Lock recall — FDA Z-2390-2016

It was reported that the secondary locking mechanism, and its corresponding components, disassociated from the Specialty Triathlon Tibial Alignment Handle…

Class IIclassification
Z-2390-2016official record ID
Aug 17, 2016report date

Recall record at a glance

Official recordZ-2390-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmStryker Howmedica Osteonics Corp.
Report dateAug 17, 2016
Recall initiation dateJun 23, 2016
Quantity47 units
LocationMahwah, NJ, United States

Official product description

Stryker Orthopaedics Specialty Triathlon Tibial Alignment Handle with Secondary Lock, Non-Sterile, Used to complement or replace the standard instruments used during the implantation of hip, knee, trauma and upper extremities implants

Why the product was recalled

It was reported that the secondary locking mechanism, and its corresponding components, disassociated from the Specialty Triathlon Tibial Alignment Handle with Secondary Lock Assembly during surgery. Upon further investigation, it was discovered that the weld, which was intended to hold the secondary locking mechanism together, did not meet the weld size specified on the engineering drawing.

Distribution information

Nationwide Distribution to FL, IN, LA, NC, PA, TN and UT

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records