FDA recall record

Stryker Orthopaedics Modular Handle Triathlon Instruments, Non-Sterile, REF 6541-4-808, Lot recall — FDA Z-2377-2016

Stryker Orthopaedic is voluntarily recalling the Triathlon Modular Handle because analysis revealed that the press fit specifications between the dowel…

Class IIclassification
Z-2377-2016official record ID
Aug 17, 2016report date

Recall record at a glance

Official recordZ-2377-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmStryker Howmedica Osteonics Corp.
Report dateAug 17, 2016
Recall initiation dateJun 23, 2016
Quantity12,469 units
LocationMahwah, NJ, United States

Official product description

Stryker Orthopaedics Modular Handle Triathlon Instruments, Non-Sterile, REF 6541-4-808, Lot 123, NDC M546654148080S Product Usage: Is a reusable instrument that is part of the Triathlon Primary total knee arthroplasty (TKA) system as well as the Triathlon Partial Knee Replacement (PKR) system. The use of the handles in Primary TKA is optional and may be used to: (1) Assist in stabilizing the Triathlon Femoral Sizer, (2) Aid in stabilization during use and removal of the Triathlon 4:1 Cutting Block and (3) Aid in assembly of the Anterior MIS Distal Resection Guide to the Anterior Skim Cut Guide. Per the Triathlon PKR surgical protocol, the handle is required for a PKR surgery and is used to: (1) Assemble and disassemble the Femoral Trial Drill Guide into the Femoral Trial and (2) Remove the Peg Trial from the Femoral Trial.

Why the product was recalled

Stryker Orthopaedic is voluntarily recalling the Triathlon Modular Handle because analysis revealed that the press fit specifications between the dowel pin and the mating hole in the shaft were not being achieved potentially resulting in disassociation from the instrument.

Distribution information

Worldwide Distribution - US Nationwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records