Recall record at a glance
| Official record | Z-0370-2025 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Stryker Corporation |
| Report date | Nov 20, 2024 |
| Recall initiation date | Oct 4, 2024 |
| Quantity | 403,730 eaches |
| Location | Portage, MI, United States |
Official product description
Stryker Neptune SafeAir Smoke Evacuation Pencil Coated Push-Button, Bulk Non-sterile - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect. Catalog Number: 0703-046-002
Why the product was recalled
There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
Distribution information
Worldwide distribution - US Nationwide and the countries of Canada, Japan, Australia.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.