FDA recall record

STERIS 5085 and 5085 SRT Surgical Tables, Product Usage: The 5085 is a general surgical recall — FDA Z-2426-2012

STERIS has learned from Customer feedback and field service experience that damage from external impact to the table hydraulic column can cause the…

Class IIclassification
Z-2426-2012official record ID
Sep 26, 2012report date

Recall record at a glance

Official recordZ-2426-2012
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSteris Corporation
Report dateSep 26, 2012
Recall initiation dateApr 6, 2012
Quantity379 tables
LocationMentor, OH, United States

Official product description

STERIS 5085 and 5085 SRT Surgical Tables, Product Usage: The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation. It accommodates all general surgical procedures including but not limited to, cardiac and vascular, endoscopic, gynecology, urology, nephrectomy, neurology, ophthalmologic, orthopedics and other procedures requiring intraoperative fluoroscopic C-arm imaging and also supports laparoscopic surgical technique for the largest surgical patients. The 5085SRT table is a variation of the 5085 table that enables patient transport on hard level surfaces within the surgical suite (from pre-operative areas to the operating room and again from the operating room to post operative recovery).

Why the product was recalled

STERIS has learned from Customer feedback and field service experience that damage from external impact to the table hydraulic column can cause the stainless steel shroud sections covering the table hydraulic column to become misaligned; if the damage is severe, the up/down movement of the table may be impeded. In addition, the table hand control may indicate that the table floor locks are unlock

Distribution information

Worldwide Distribution - US Nationwide including the states of: AL, AR, CA, CO, CT, FL, HI, ID, IL, IN, IA, KY, LA, MA, MD, ME, MI, MN, MO, MS, NV, NYH, NC, ND, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WI & WV and countries of: Australia, Canada, China, India, Korea, Mexico, Panama and Thailand.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.