FDA recall record

Sterilmed Reprocessed Imaging Catheter ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR FOR recall — FDA Z-0574-2026

Affected catheters were reprocessed beyond validated cycles. This may compromise device sterility or mechanical integrity of the device and can lead to…

Class IIclassification
Z-0574-2026official record ID
Nov 26, 2025report date

Recall record at a glance

Official recordZ-0574-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmSterilmed, Inc.
Report dateNov 26, 2025
Recall initiation dateOct 8, 2025
Quantity4
LocationPlymouth, MN, United States

Official product description

Brand Name: Sterilmed Reprocessed Imaging Catheter Product Name: ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR FOR USE WITH SIEMENS SYSTEM Model/Catalog Number: ACU10135936 Software Version: N/A Product Description: ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR FOR USE WITH SIEMENS SYSTEM Component: N/A

Why the product was recalled

Affected catheters were reprocessed beyond validated cycles. This may compromise device sterility or mechanical integrity of the device and can lead to severe adverse events such as infection, vascular injury or embolism.

Distribution information

United States Only: Kentucky, Maryland, Massachusetts, Mississippi, New Jersey, Pennsylvania, Texas, Utah

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Telephone.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.