Recall record at a glance
| Official record | Z-1757-2017 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | ConMed Corporation |
| Report date | Apr 19, 2017 |
| Recall initiation date | Feb 28, 2017 |
| Quantity | N/A |
| Location | Utica, NY, United States |
Official product description
Sterile Handpiece Sheaths, P/N 7-796-19, included in the Accessory Packages contained in the HYFRECATOR 2000 Electrosurgical Unit (ESU) cartons, 115 VOLTS AC, REF/Catalog Number 7-900-115, Rx Only . There is no problem with the Hyfrecator 2000 Electrosurgical Units. -- ASSEMBLED IN MEXICO --- CONMED CORPORATION 525 FRENCH ROAD UTICA, NY 13502 The Hyfrecator 2000 Electrosurgical Unit is indicated for use in conjunction with an electrosurgical accessory handpiece for delivery of high frequency electrosurgical current through an accessory electrode for cutting and coagulation at the operative site. The sterile sheaths are intended to cover the electrosurgical handpiece during those typical electrosurgical procedures and reduce the effort necessary to re-sterilize the handpiece between procedures.
Why the product was recalled
For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were sterile.
Distribution information
Worldwide Distribution - US (nationwide) Internationally to Australia, Canada, Germany, Italy, UK, Great Britain, and Ireland.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.