NOVADAQ TECHNOLOGIES Sterile Drape, SK6000 PAQ drape pouch recall — FDA Z-2767-2016

Novadaq is informing SK6000 PAQ users that a pinhole through the drape pouch film material was identified in some SK6000 PAQs during shipping validation…

Class IIclassification
Z-2767-2016official record ID
Sep 14, 2016report date

Recall record at a glance

Official recordZ-2767-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmNOVADAQ TECHNOLOGIES Inc.
Report dateSep 14, 2016
Recall initiation dateAug 18, 2016
Quantity20 units
LocationBurnaby, Canada

Official product description

Sterile Drape, SK6000 PAQ drape pouch (for use with the SPY Fluorescent System (SPY 2000/2001)) The SPY Fluorescent Imaging System is intended to provide fluorescent images for the visual assessment of blood flow in vessels and related tissue perfusion during cardiovascular surgical procedures. For use in intraoperative visual assessment of the coronary vasculature and bypass grafts during coronary artery bypass (BABG) surgery.

Why the product was recalled

Novadaq is informing SK6000 PAQ users that a pinhole through the drape pouch film material was identified in some SK6000 PAQs during shipping validation testing.

Distribution information

US Distribution in states of: FL, MI, and VA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.