Recall record at a glance
| Official record | Z-1455-2025 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | CooperVision, Inc. |
| Report date | Apr 2, 2025 |
| Recall initiation date | Feb 27, 2025 |
| Quantity | 33990 lenses (17010 US, 16980 OUS) |
| Location | West Henrietta, NY, United States |
Official product description
Brand Names: MyDay Toric, LensCrafters One Day Premium Daily Disposable Contact Lenses For Astigmatism, Pearle Vision 1 Day Premium For Astigmatism, REVEAL 1-DAY TORIC, Vision Source Reveal Toric, Voyant 1-Day Premium Toric. Brand Names (OUS only): CADENCE DELUXE 1-DAY TORIC, DUBLAN PREMIUM OXYGEN 1 DAY TORIC, EV LINARIAL TORIC (30), EYEXPERT PURE ASTIGMATISM (30), LENTI!!AS A ASTIGMATISMO 30PK, LENTSOFT DIARIA CONFORT+ SILICONA ASTIGMATISMO, MULTILENS CONTACT ELITE SILICONE 1 DAY TORICS, OPTIONS SUPREME 1 DAY TRC (30), PRIMA + 1 DAY ASTGM (30), PROTECT PLATINUM FOR ASTG (30), SOL- LENS 1 DAY SILICONE PREMIUM TORIC, UGO LENS DAY ADVANCED TORICA, WHITE BOX STENFILCON A TORIC 07536, MEDIFLEX SUP 1D TORIC, Kirkland Signature Toric Premium Daily Disposable, Specsavers easyvision Linarial Toric. Product Name: (Stenfilcon A) Contact Lens Model/Catalog Number: MyDay Toric Product Description: 54% water content, Toric, cast moulded Silicone Hydrogel, In-monomer tinted, contact lens, for Daily Disposable Wear. Blister packaged. Component: No
Why the product was recalled
a limited number of lots were manufactured with an incorrect cylinder power.
Distribution information
Domestic: CA, CO, FL, IL, KY, MI, NC, NE, NJ, NY, OH, OR, UT, VA, WA. International: Australia, Belgium, Canada, Czech Republic, Denmark, France, Germany, Ireland, Italy, Luxembourg, Netherlands, New Zealand, Norway, Poland, Portugal, Spain, Sweden, Switzerland, and United Kingdom.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.