FDA recall record

Nova Biomedical StatStrip Glucose Hospital Meter System recall — FDA Z-3128-2026

A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a…

Class Iclassification
Z-3128-2026official record ID
Sep 23, 2026report date

Recall record at a glance

Official recordZ-3128-2026
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmNova Biomedical Corporation
Report dateSep 23, 2026
Recall initiation dateJul 29, 2026
Quantity20,616 units
LocationWaltham, MA, United States

Official product description

StatStrip Glucose Hospital Meter System. Model No: 63685. Glucose test system.

Why the product was recalled

A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a new barcode is scanned, resulting in the prior barcode information remaining displayed on the screen for the current test. When this defect occurs and the operator accepts the incorrectly displayed patient information without recognizing the mismatch, the current patient's glucose result will be incorrectly assigned to the previous patient's medical record and transmitted to the data management system (DMS), and the current patient's medical record will not receive a result, leaving their glucose status unrecorded and their clinical management without the information needed to guide treatment decisions. The previous patient, whose medical record now contains a glucose result that does not belong to them, may receive inappropriate clinical intervention based on that erroneous result, particularly in critically ill patients, neonates, and infants who cannot communicate symptoms or self-treat. The defect also affects QC barcode scanning: if a QC scan triggers the same retention behavior, QC results may be attributed to an incorrect identifier or QC tests may fail without the user recognizing the underlying cause. This defect does not affect the analytical accuracy of glucose measurements.

Distribution information

Worldwide distribution: US (nationwide) and OUS (foreign) countries of: Austria, Belgium, Bolivia, Brazil, Canada, Cayman Islands, Chile, Croatia, Dominican Rep., Finland, France, Germany, Ireland, Israel, Italy, Netherlands, New Zealand, South Korea, Switzerland, Thailand, Turkey, United Kingdom.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.