FDA recall record

Ventlab STAT-Check Resuscitation Bag Manual Emergency Resuscitation Bag with Carbon dioxide recall — FDA Z-1880-2016

Stat-check CO2 indicators are non-functional. The indicator is yellow when not in use and should be purple when not in use. The yellow indicates a 5% CO2…

Class IIclassification
Z-1880-2016official record ID
Jun 8, 2016report date

Recall record at a glance

Official recordZ-1880-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmVentlab LLC
Report dateJun 8, 2016
Recall initiation dateMar 15, 2016
Quantity2430
LocationGrand Rapids, MI, United States

Official product description

STAT-Check Resuscitation Bag Manual Emergency Resuscitation Bag with Carbon dioxide gas analyzer Intended for use in various clinical situations to provide respiratory support in the presence of reversible apnea commonly associated with respiratory arrest. The device should only be used by personnel trained in CPR procedures. The ACE/STAT-Check II is intended for use as an adjunct assessment tool for verification of successful ventilation / intubation in patients > 2.2 kilograms (5lbs.).

Why the product was recalled

Stat-check CO2 indicators are non-functional. The indicator is yellow when not in use and should be purple when not in use. The yellow indicates a 5% CO2 level, cardiac massage and ventiliation being performed correctly. The color will not change during use and will not be an indicator of CO2 level or improper performance of cardiac message and ventilation.

Distribution information

Nationwide distribution to AZ, MS, NY, MO, FL, WI, TX, FL, and CA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.