FDA recall record

Spiegelberg Brain Pressure Monitoring System Probe 3PN, REF: SNDI3.1.531FV534P and Probe recall — FDA Z-2888-2016

In the vast majority of products specified, a shrinkage of the air chamber during sterilization has occurred. Because of these false high readings were…

Class IIclassification
Z-2888-2016official record ID
Oct 5, 2016report date

Recall record at a glance

Official recordZ-2888-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSpiegelberg Gmbh & Co. KG
Report dateOct 5, 2016
Recall initiation dateDec 29, 2009
Quantity22 units
LocationHamburg, N/A, Germany

Official product description

Spiegelberg Brain Pressure Monitoring System Probe 3PN, REF: SNDI3.1.531FV534P and Probe 3PS, REF: SNDI3.1.63/FV535P Product Usage: lntracranial pressure measuring probes are used to measure intracranial pressure (intracranial pressure, ICP), so the pressure in the head.

Why the product was recalled

In the vast majority of products specified, a shrinkage of the air chamber during sterilization has occurred. Because of these false high readings were displayed in the lower measuring range from 0 to 20 mmHg. The observed measurement error was particularly at the lowest end of the range (to 0 mmHg) up to 8 mmHg.

Distribution information

Nationwide Distribution to PA only

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.