Recall record at a glance
| Official record | Z-3106-2017 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class I |
| FDA record status | Terminated |
| Recalling firm | Spectranetics Corporation |
| Report date | Oct 4, 2017 |
| Recall initiation date | Aug 7, 2017 |
| Quantity | 2,356 units |
| Location | Colorado Springs, CO, United States |
Official product description
Spectranetics Bridge Occlusion Balloon catheter, Sheath Compatibility 12F, Guidewire Compatibility 0.035in, Working Length 90cm, Balloon Outer Diameter 31mm, Balloon Length 80mm, Rx Only. Product Usage: It is a device used to occlude the vessel and prevent blood loss as the physician preps for surgery resulting from an SVC tear complication of a lead extraction case. It is used for temporary vessel occlusion of the superior vena cava in applications including perioperative occlusion and emergency control of hemorrhage.
Why the product was recalled
Possible inability to pass the guidewire through the Bridge device, therefore, resulting in a potential delay of treatment.
Distribution information
Worldwide Distribution: US (nationwide) and countries of: Australia, Austria, Belgium, Cyprus, Denmark, Finland, France, Germany, Hungary, Iceland, Ireland, Israel, Italy, Kuwait, Netherlands, Norway, Poland, South Africa, Sweden, Switzerland, and United Kingdom.
How to read the classification
FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.