FDA recall record

DePuy Orthopaedics Specialist 2 IM Rod, Non-Sterile and associated Surgical Techniques recall — FDA Z-0971-2013

Potential for the IM rod to break, leaving fragments in the patient. Surgical Techniques guidance has been updated until design changes can be made to…

Class IIclassification
Z-0971-2013official record ID
Mar 27, 2013report date

Recall record at a glance

Official recordZ-0971-2013
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmDePuy Orthopaedics, Inc.
Report dateMar 27, 2013
Recall initiation dateFeb 5, 2013
Quantity4000; 2600 in USA
LocationWarsaw, IN, United States

Official product description

Specialist 2 IM Rod, Non-Sterile and associated Surgical Techniques: Sigma HP Instruments Classic Surgical Technique,0612-89-510 Sigma HP Instruments Balanced Surgical Technique, 0612-88-510 Sigma HP Instruments Fixed Reference Surgical Technique, 0612-87-510 PFC Sigma Knee Systems Primary Cruciate-Retaining & Cruciate-Substituting Procedure Surgical Technique,SP2-007 Product Usage: The SP2 IM Rod is used in both primary and revision Sigma knee procedures to align the femoral locating device and distal femoral cutting block. It is also used with the IM tibial resection. This rod is included in both the SP2 and HP instrument kits.

Why the product was recalled

Potential for the IM rod to break, leaving fragments in the patient. Surgical Techniques guidance has been updated until design changes can be made to reduce likelihood of potential failure.

Distribution information

Worldwide Distribution - USA Nationwide.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.