FDA recall record

Spacelabs qube Compact Monitor, Model 91390 recall — FDA Z-1608-2015

The input power can fail even though the power source at the wall has not been interrupted. In addition, there is the potential for the qube Compact…

Class IIclassification
Z-1608-2015official record ID
May 20, 2015report date

Recall record at a glance

Official recordZ-1608-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSpacelabs Healthcare Inc.
Report dateMay 20, 2015
Recall initiation dateMay 8, 2015
Quantity5 units distributed in the US
LocationSnoqualmie, WA, United States

Official product description

Spacelabs qube Compact Monitor, Model 91390. The monitor is functioning as either bedside or central monitors; passively displays data generated by Spacelabs Healthcare parameter modules, Flexports interfaces, and other SDLC based products in the form of waveform and numeric displays, trends and alarms.

Why the product was recalled

The input power can fail even though the power source at the wall has not been interrupted. In addition, there is the potential for the qube Compact Monitor to fail to turn on. This is a retrospective recall related to a recall initiated in 2012 (Z-0403-2013).

Distribution information

distributed in North Carolina, Texas, and Wyoming

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.