FDA recall record

Del Mar Reynolds Medical SpaceLabs Arkon Anesthesia Delivery System with Version 2.0 recall — FDA Z-1442-2014

The System may go into a controlled failed state if a series of parameters on the user interface are activated in a specific sequence.

Class Iclassification
Z-1442-2014official record ID
Apr 23, 2014report date

Recall record at a glance

Official recordZ-1442-2014
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmDel Mar Reynolds Medical, Ltd.
Report dateApr 23, 2014
Recall initiation dateMar 10, 2014
Quantity16 units distributed in the US
LocationHertford, N/A, United Kingdom

Official product description

SpaceLabs Arkon Anesthesia Delivery System with Version 2.0 Software installed. This system includes a ventilator.

Why the product was recalled

The System may go into a controlled failed state if a series of parameters on the user interface are activated in a specific sequence.

Distribution information

USA Nationwide Distribution in the states of North Carolina and South Carolina

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.