FDA recall record

Sorin Platinium CRT-D 1741 DF-4, Sterile EO recall — FDA Z-2825-2018

Hardware configuration may lead to overconsumption, followed by loss of pacing and sensing capabilities. The device then cannot identify an arrhythmia…

Class IIclassification
Z-2825-2018official record ID
Aug 22, 2018report date

Recall record at a glance

Official recordZ-2825-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSorin CRM SAS
Report dateAug 22, 2018
Recall initiation dateJul 9, 2018
Quantity10
LocationClamart Cedex, France

Official product description

Sorin Platinium CRT-D 1741 DF-4, Sterile EO. Implantable Cardiac Defibrillator and Resynchronization Therapy Defibrillator

Why the product was recalled

Hardware configuration may lead to overconsumption, followed by loss of pacing and sensing capabilities. The device then cannot identify an arrhythmia that would require defibrillation shock therapy.

Distribution information

US distribution to: NJ, AZ, MS, OK, LA, PA, FL, MI, IL, and MS

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.