FDA recall record

LivaNova USA SORIN Heater-Cooler System 3T 120V~/60Hz REF 16-02-85 Description of product recall — FDA Z-0220-2020

The firm has become aware that due to a chemical reaction, the disinfectant/water preservative used to disinfect the device to inhibit/limit the growth…

Class IIclassification
Z-0220-2020official record ID
Nov 6, 2019report date

Recall record at a glance

Official recordZ-0220-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmLivaNova USA Inc.
Report dateNov 6, 2019
Recall initiation dateNov 1, 2018
Quantity8671
LocationArvada, CO, United States

Official product description

SORIN Heater-Cooler System 3T 120V~/60Hz REF 16-02-85 Description of product: The St¿ckert Heater-Cooler System is an independent (i.e. independent of the water supply) 3-circuit-heating/cooling system. The three separate water circuits have been developed to serve as: " Two circuits for the patient supplied by one tank (for the heating/cooling blanket and the heat-exchanger associated with the oxygenator), common temperatures ranging from 2 degrees Celsius to 41 degrees Celsius, tank volume 6 liters. " One interchangeable heating/cooling circuit for cardioplegia heat-exchanger, temperatures ranging from 2 degrees Celsius to 10 degrees Celsius (cooling tank) and/or from 15 degrees Celsius to 41 degrees Celsius (heating tank). If required, the circuits for the patient and the circuit for cardioplegia can be switched off separately, in order to increase the activated functional group s heating and/or cooling performance. Product Usage: The Stockert Heater-Cooler System 3T is intended for use with a Stockert/Sorin S3/S5/C5 heart-lung machine and/or any other heart-lung machine featuring a separate temperature control for extracorporeal perfusion of durations of up to 6 hours.

Why the product was recalled

The firm has become aware that due to a chemical reaction, the disinfectant/water preservative used to disinfect the device to inhibit/limit the growth micro-organisms, may be rendered ineffective. In addition, the firm is releasing a design upgrade to reduce the risk of potential emission of aerosols from the 3T.

Distribution information

Worldwide distribution - US Nationwide distribution including sate of PR and countries of Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Georgia, Germany, Greece, Guatemala, Guyana, Hong Kong, Hungary, India, Indonesia, Iran (Islamic Republic of), Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea (the Republic of), Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia (the former Yugoslav Republic of), Malaysia, Maldives, Malta, Mauritius, Mexico, Monaco, Mongolia, Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines (the), Poland, Portugal, Qatar, Romania, Russian Federation (the), Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan (Province of China), Thailand, Trinidad and Tobago, Tunisia, Turkmenistan, Turkey, Ukraine, United Arab Emirates (the), United Kingdom, Uruguay, Uzbekistan, Vietnam.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.