Recall record at a glance
| Official record | Z-0692-2017 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Siemens Medical Solutions USA, Inc. |
| Report date | Dec 14, 2016 |
| Recall initiation date | Oct 20, 2016 |
| Quantity | 48 units distributed in U.S. |
| Location | Malvern, PA, United States |
Official product description
SOMATOM Force, System x-ray, tomography, computed
Why the product was recalled
Siemens is providing software update version VA50A_SP3 to address the software bugs that were identified through normal field monitoring and the Global Complaint Handling Process. Correction for the problems are as follows: 1. Correction to volumetric misrepresentations of high contrast objects when using ADMIRE. 2. Correction to highly sporadic scan aborts due to temporarily tube currents at 0mA. 3. Correction to missing Microsoft Hotfixes (MS16-001, MS15-135, MS15-088, MS15-048). 4. Correction to highly sporadic image artifacts caused by error signals from high voltage chain at extremely high tube currents. 5. Correction to highly sporadic incomplete image reconstruction due to temporary slow hard disk write speed. 6. Correction to highly sporadic scan aborts due to temporarily high disk load of system disk. 7. Correction to highly sporadic image artifacts generated by scatter correction in case of tube arcings.
Distribution information
Nationwide Distribution.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.