FDA recall record

Olympus Corporation of the Americas SOLTIVE SuperPulsed Laser Fibers Model Numbers recall — FDA Z-0748-2024

Olympus identified inconsistencies in the Instructions for Use regarding cleaning the fiber tip and fiber cleaving and stripping instructions. Following…

Class IIclassification
Z-0748-2024official record ID
Jan 24, 2024report date

Recall record at a glance

Official recordZ-0748-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmOlympus Corporation of the Americas
Report dateJan 24, 2024
Recall initiation dateDec 1, 2023
Quantity82,200 total units (5 per box)
LocationCenter Valley, PA, United States

Official product description

SOLTIVE SuperPulsed Laser Fibers Model Numbers TFL-FBX150S, TFL-FBX200S, TFL-FBX365S, TFL-FBX550S, TFL-FBX940S, TFL-FBX150BS, TFL-FBX200BS

Why the product was recalled

Olympus identified inconsistencies in the Instructions for Use regarding cleaning the fiber tip and fiber cleaving and stripping instructions. Following the current instructions presents the potential for the fiber to break during use, which may cause burns, procedure delays, prolonged surgery, and foreign body in patient.

Distribution information

Worldwide distribution - US Nationwide and the countries of Canada, Europe, Hong Kong, Australia, Japan, Singapore, and Latin America.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.