FDA recall record

Quidel Sofia 2 Analyzer Catalog Number 20299 Sofia 2 is a bench top analyzer intended to be recall — FDA Z-1183-2018

The calibration for the affected batch of analyzers was incorrectly set during the manufacturing process which may lead to a failure when calibration is…

Class IIclassification
Z-1183-2018official record ID
Apr 4, 2018report date

Recall record at a glance

Official recordZ-1183-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmQuidel Corporation
Report dateApr 4, 2018
Recall initiation dateDec 15, 2017
Quantity482 units
LocationSan Diego, CA, United States

Official product description

Sofia 2 Analyzer Catalog Number 20299 Sofia 2 is a bench top analyzer intended to be used with Cassette-based immunofluorescent in vitro diagnostic assays . Sofia 2 is intended for professional and laboratory use.

Why the product was recalled

The calibration for the affected batch of analyzers was incorrectly set during the manufacturing process which may lead to a failure when calibration is first performed. There may also be e remote possibility that incorrect qualitative test results for patient samples near the assay cutoff may be reported.

Distribution information

US Distribution

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.