Recall record at a glance
| Official record | Z-2331-2020 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Smith & Nephew, Inc. |
| Report date | Jun 17, 2020 |
| Recall initiation date | Apr 22, 2020 |
| Quantity | 1267 devices |
| Location | Memphis, TN, United States |
Official product description
smith&nephew R3 NO HOLE HEMISPHERICAL STIKTITE COATED SHELL, manufactured in the following sizes: Size REF/product number 42MM 71331842 48MM 71331848 50MM 71331850 52MM 71331852 54MM 71331854 56MM 71331856 58MM 71331858 60MM 71331860 Orthopedic implant component.
Why the product was recalled
A manufacturing error resulted in out of specification R3 Acetabular Shells
Distribution information
Worldwide
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.