Recall record at a glance
| Official record | Z-2250-2017 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Circadiance LLC |
| Report date | Jun 14, 2017 |
| Recall initiation date | May 1, 2017 |
| Quantity | 1,242 units |
| Location | Export, PA, United States |
Official product description
SmartMonitor 2 PS/PSL, Monitor, Apnea, Facility USE. Intended for use in the continuous monitoring of respiration, heart rate, and SP02 of infant, pediatric and adult patients.
Why the product was recalled
Circadiance has determined that it is possible for certain Smart Monitor 2PS/PSL monitors to exhibit intermittent operation of the nurse call feature. The firm has updated the device to increase the "Nurse Call" feature to ensure monitor alarms are transmitted to compatible nurse call systems.
Distribution information
Nationwide Distribution.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.