Single Use Ligating Device Olympus HX-400U-30 recall — FDA Z-0570-2026

Potential that the ligation loop was unable to release or detach as expected during use, causing the loop to become unintentionally fixed around patient…

Class Iclassification
Z-0570-2026official record ID
Dec 3, 2025report date

Recall record at a glance

Official recordZ-0570-2026
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmOlympus Corporation of the Americas
Report dateDec 3, 2025
Recall initiation dateOct 30, 2025
Quantity7803 units
LocationCenter Valley, PA, United States

Official product description

Brand Name: Single Use Ligating Device Product Name: Olympus HX-400U-30 Model/Catalog Number: HX-400U-30 Used with an Olympus endoscope to deliver a nylon loop snare.

Why the product was recalled

Potential that the ligation loop was unable to release or detach as expected during use, causing the loop to become unintentionally fixed around patient anatomy.

Distribution information

US-wide distribution

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.