FDA recall record

Covidien Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT recall — FDA Z-1457-2024

Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue…

Class IIclassification
Z-1457-2024official record ID
Apr 10, 2024report date

Recall record at a glance

Official recordZ-1457-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmCovidien, LP
Report dateApr 10, 2024
Recall initiation dateFeb 14, 2024
Quantity30 Units
LocationNorth Haven, CT, United States

Official product description

Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT

Why the product was recalled

Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.

Distribution information

Worldwide distribution - US Nationwide and the countries of France, Hong Kong, Israel and Japan.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.