Recall record at a glance
| Official record | Z-2921-2020 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Siemens Medical Solutions USA, Inc. |
| Report date | Sep 9, 2020 |
| Recall initiation date | Jul 24, 2020 |
| Quantity | 110 units US |
| Location | Malvern, PA, United States |
Official product description
Siemens Sensis Vibe Hemo system in combination with the MicroPodTM EtCO2 module-diagnostic and administrative tool supporting hemodynamic catheterization and/or electrophysiology studies, for cardiac as well as interventional Radiology.
Why the product was recalled
The connector linking the EtCO2 module to the HemoBox may break and become damaged, if the connector is re-attached to the HemoBox in this damaged state, under certain incorrect pin alignment conditions, the HemoBox may stop functioning and all vital signs are lost
Distribution information
US Nationwide distribution.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.